Shop By Categories
© 2023.Insuzo. All Rights Reserved.
Electrical and Electronic
Enviornmental
General Laboratory | Analytical
Imaging Instruments
Mechanical and Quality Control
Time, Speed and Dimension
Calibrators
Shop By Industry
Item(s) Added To Your Bag
MECO PHA 5850B Power & Harmonics Analyzer
MECO PHA 5850B Power & Harmonics Analyzer
(Based On 10 Reviews)
MECO Power and Harmonics Analyser Model PHA-5850 can Analyze, Measure, Monitor &Data Log values of Power Quality & Consumption (Energy).
Capable of analyzing IT standby power consumption to the maximum demand of factory. It comes with a user friendly application software that increases
the utility & performance of this instrument. The analyzer is ideal for an any Engineer / Inspector for carrying out Periodic Visits, Maintainance of Plant, Vigilance
checks, Surveys and Energy Audits for checking at Industrial and Consumers end.
Request Callback
Order In Bulk
Generate Quote
₹ 2,62,881.00
(GST inclusive price) i₹2,62,881.00
( - 0% OFF )
Description
What's in the box?
|
Rating & Reviews
Rating
General Specification | |
Power | Eight x 1.5V AA Batteries |
External DC Input | Power supply adapter 12 Volts. DC |
Display | Dot Matrix LCD (240x128) with backlight |
LCD Update Rate | 1 time / second |
Power Consumption | 140mA (approx.) |
No. Of Samples | 1024 samples / period |
Data Logging Files | 85 |
Max. File Capacity | 17474 records (3P4W, 3P3W) 26210 records (1P3W) 52420 records (1P2W) 4096 records (50 Harmonics / record) 2 to 3000 seconds for data logging |
Sampling Time | OL |
Operating Condition | -10°C to 50°C < 85% RH |
Storage Condition | -20°C to 60°C < 75% RH |
Dimensions | 257 x 155 x 57 mm |
Weight | 1160 g (Batteries included) |
Model : PHA-5850A (100A) (Overload Protection AC 200A) | ||
Range | Resolution | Accuracy of Readings |
0.04 - 1A | 0.1mA / 1mA | ± 0.5% ± 0.05A |
0.4 - 10A | 0.001A / 0.01A | ± 0.5% ± 0.05A |
4 - 100A | 0.01A / 0.1A | ± 1.0% ± 0.5A |
Model : PHA-5850B (1000A) (Overload Protection AC 2000A) | ||
Range | Resolution | Accuracy of Readings |
10.00A | 0.001A / 0.01A | |
5A - 100.0A | 0.01A / 0.1A | ± 0.5% ± 0.5A |
50A - 1000.0A | 0.1A / 1A | ± 0.5% ± 5A |
Model : PHA-5850C (3000A) (Overload Protection AC 3000A) | ||
Range | Resolution | Accuracy of Readings |
10.0 - 300.0A | 0.01A / 0.1A | ± 1% of range |
300.0 - 3000A | 0.1A / 1A |
Model : PHA-5850D (1200A) (Overload Protection AC 1200A) | ||
Range | Resolution | Accuracy of Readings |
6.0 - 120.0A | 0.01A / 0.1A | ± 1% of range |
300.0 - 3000A | 0.1A / 1A |
Harmonic of AC Voltage in Percentage | ||
Range | Resolution | Accuracy of Readings |
1 - 10th | 0.1% | ± 2% |
11 - 20th | ± 4% of reading ± 2.0% | |
50 - 99th | ± 6% of reading ± 2.0% |
Harmonic of AC Current in Percentage Model : PHA-5850A (100A) | ||
Range | Resolution | Accuracy of Readings |
1 - 10th | 0.1% | ± 0.2% of reading ± 1% |
11 - 20th | ± 2% of reading ± 1% | |
21 - 50th (A range) | ± 5% of reading ± 1% | |
21 - 50th (mA range) | ± 10% of reading ± 1% | |
51 - 99th | ± 35% of reading ± 1% |
Model : PHA-5850B (1000A) | ||
Range | Resolution | Accuracy of Readings |
1 - 20th | 0.1% | ±2% |
21 - 49th | ± 4% of reading ± 2.0% | |
50 - 99th | ± 6% of reading ± 2.0% |
Model : PHA-5850C (3000A) & PHA-5850D (1200A) | ||
Range | Resolution | Accuracy of Readings |
1 - 20th | 0.1% | ±2 % |
21 - 49th | ± 6% | |
50 - 99th | ± 10% |
Model : PHA-5850A (100A) | ||
Range | Resolution | Accuracy of Readings |
5.0 - 999.9W | 0.1W | ± 1% ± 0.8W |
1.000 - 9.999KW | 0.001KW | ± 1% ± 8W |
10.00 - 99.99KW | 0.01KW | ± 1% ± 80W |
100.0 - 999.9KW | 0.1KW | ± 1% ± 0.8KW |
1000 - 9999KW | 1KW | ± 1% ± 8KW |
Model : PHA-5850B (100A) | ||
Range | Resolution | Accuracy of Readings |
5.0 - 999.9W | 0.1W | ± 1% ± 0.8W |
1.000 - 9.999KW | 0.001KW | ± 1% ± 8W |
10.00 - 99.99KW | 0.01KW | ± 1% ± 80W |
100.0 - 999.9KW | 0.1KW | ± 1% ± 0.8KW |
1000 - 9999KW | 1KW | ± 1% ± 8KW |
0.000 - 9.999MW | 0.001MW | ± 1% ± 80KW |
Model : PHA-5850C (3000A) Model : PHA-5850D (1200A) | |||
Range | Resolution | Accuracy of Readings | |
10.0 - 999.9W | 0.1W | >20V & >30A | <20V or <30A |
1.000 - 9.999KW | 0.001KW | ± 1% of range | ± 2% of range |
10.00 - 99.99KW | 0.01KW | ||
100.0 - 999.9KW | 0.1KW | ||
1000 - 9999KW | 1KW |
AC Voltage (50Hz or 60H | ||
Range | Resolution | Accuracy of Readings |
20.0V - 500.0V (Phase to Neutral) | 0.1V | ± 0.5% ± 5dgts |
Power Factor (PF) Model : PHA-5850A (100A) & PHA-5850B (1000A) | ||
Range | Resolution | Accuracy of Readings |
0.00 - 1.00 | 0.1V | ± 0.04 |
Model : PHA-5850C (3000A) & PHA-5850D (1200A) | |||
Range | Resolution | Accuracy of Readings | |
0.000 - 1.000 | 0.001 | >20V & >30A | <20V or <30A |
± 0.04 | ± 0.1 |
Rating & Reviews
Rating
Similar Products
More From Brands
Request Callback
We will get in touch with you
Enter your requirement details!
We will get in touch with you.
Share This Product
Reach Us
The Office Address
305 P.P. Trade Center, Netaji Subash Place, Pitampura, Delhi - 110034
For Support
8920884830
Send Us Email
info@insuzo.com
Copyright © 2024. Insuzo
All Rights Reserved.